INSTALLATION MANUALAPRIL 2010PLEASE READ ENTIRE MANUAL PRIOR TO INSTALLING THIS PRODUCT.Con TROLL PRO
10 Mounting the Controller3.2 MOUNTING OPTIONS3.2.1 Controller Dimensions• Figure 4 calls out the controller dimensions with the lid closed.• Figure 5
Mounting the Controller 11FIGURE 6. Side view dimensional drawings3.2.2 Pole Mounting1. Remove the four nuts and four screws from the mounting kit.2.
12 Mounting the ControllerFIGURE 8. Attaching the mounting bracketsFIGURE 9. Back and side view of pole-mounted controller3.2.3 Wall Mounting Using Mo
Mounting the Controller 133.2.4 Wall Mounting with User-supplied Screws and HardwareYou can supply your own mounting hardware and attach the controlle
14 Mounting the Controller3.2.5 DIN Rail Mounting the EnclosureEach rail mount clip is secured to the enclosure with a supplied #6 screw.1. With the e
Mounting the Controller 155. Thread sensor or electrical cables through the dome connectors for later wiring by a qualified electrical technician. Tig
Electrical Connections 17Chapter 4 Electrical ConnectionsDangerOnly properly trained and qualified personnel should install the Con TROLL PRO instrume
18 Electrical Connections• For hard-wired locations using conduit, a 100-240 VAC or 9-36 VDC source with over current/disconnect protection will be pr
Electrical Connections 194.7 AC ELECTRICAL CONNECTIONSDangerMake sure that power to the instrument is disconnected before making any wiring connection
2Copyright © 2009-2010 by In-Situ Inc. All rights reserved.This document contains proprietary information which is protected by copyright. No part of
20 Electrical ConnectionsFIGURE 19. AC circuit board and terminal connections1. Disconnect power to the instrument.2. Use a Phillips screwdriver to re
Electrical Connections 214.8 RELAY CONNECTIONSDangerMake sure that power to the instrument is disconnected before making any wiring connections.Danger
Input/Output Connections 23Chapter 5 Input/Output ConnectionsThe Input/Output board is shown in Figure 20. It is located inside the enclosure on the l
24 Input/Output Connections5.1 PLC CONNECTIONSDangerMake sure that power to the instrument is disconnected before making any wiring connections.CONNEX
Input/Output Connections 25CONNEXIONS POUR LE RACORD DES SONDESDangerDébrancher toute alimentation à l'appareil avant de connecter les fils.1. Di
26 Input/Output Connections3. Connect the stripped and tinned, twisted-pair shielded, 24-12 AWG, 4-20 mA wires to the terminal strip and Protective Ea
Optional Battery Box 27Chapter 6 Optional Battery Box 6.1 OVERVIEW OF BATTERY BOX KITThe optional external battery box kit is designed to power the D
28 Optional Battery Box6.4 CONNECT THE EXTERNAL BATTERY ENCLOSURE TO THE CON TROLL PRO ENCLOSURE1. Route the cable from the External Battery Enclosur
Maintenance 29Chapter 7 MaintenanceDangerOnly properly trained and qualified personnel should install the Con TROLL instrument described in this manua
Table of Contents 3TABLE OF CONTENTSChapter 1 Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51.1 Hazard Symbols that A
30 Maintenance1. Disconnect AC power to the instrument.2. Use a Phillips screwdriver to remove the controller lid.3. Remove the spent batteries and di
Specifications 31Chapter 8 Specifications8.1 OVERVIEWThe Con TROLL PRO controller is a process sensor controller with an easy-to-use interface that su
32 Speci f ica t ions8.3 DECLARATION OF CONFORMITY Declaration of Conformity Manufacturer: In-Situ, Inc. 221 East Lincoln Avenue Fort Collins, CO
4 Table of ContentsChapter 7 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 297.1 Cleaning the Outside of the Enclosur
Safety Information 5Chapter 1 Safety InformationPlease read this manual before unpacking or installing any part of this system.1.1 HAZARD SYMBOLS THAT
Overview 7Chapter 2 OverviewAll Con TROLL® PRO modules have the following benefits:• NEMA 4X rated• Robust keypad and display window• Easy-to-use cont
8 OverviewFIGURE 1. Internal view of DC modelFIGURE 2. Internal view of AC models
Mounting the Controller 9Chapter 3 Mounting the ControllerDangerOnly properly trained and qualified personnel should install the Con TROLL instrument