Process Optical Dissolved Oxygen Probe Operator’s Manual November 2013 0089662 Rev. 007

Page 2

7. Clean the sensor lens with the wipe provided in the kit and allow it to thoroughly dry. Inspect for scratches or dirt. 8. Remove the new

Page 3 - Table of Contents

5.0 Controller Requirements and Connections The RDO PRO Probe may be connected to a controller or logger for communication via: • Analog (4-20

Page 4 - 1.0 Introduction

5.2 Analog (4-20 mA) 3-wire Signal Color Ground/Return Black External Power (12-36 VDC) Red 4-20 mA Brown  Cable length must not ex

Page 5 - 1.3 Unpack the Probe

5.3 SDI-12 (3 wire) Signal Color Ground/Return Black External Power (9.6-16 VDC) Red RS485 (-) Green RS485 (+) Blue SDI-12 White  Ca

Page 6 - 2.1 1-point Calibration

5.4 Modbus Master with Built-in RS485 Signal Color Ground/Return Black External Power (12-36 VDC) Red RS485 (-) Green RS485 (+) Blue 

Page 7 - 2.2 2-point Calibration

5.5 Modbus Master with Built-in RS232 (Converter Required) Signal Color Ground/Return Black External Power (12 VDC, voltage limited by conve

Page 8 - 3.0 Sensor Deployment

5.5.2 Power Connections The red wire provides power for all system modes. Analog output is disabled by default. However, the 4-20 mA current lo

Page 9 - 4.0 Care and Maintenance

0049 1 R1/W2 ushort Units Id 1 = °C (default) 2 = °F 0050 1 R1 ushort Data Quality Id 0051 2 R1/W3 float Off line sentinel value (default = 0.0)

Page 10 - 4.5 Maintaining Desiccant

6.3.1 Dissolved Oxygen Concentration Equations DO concentration is internally calculated in mg/L. Conversion to other units is as follows: µg/L

Page 11 - 5.1 Wiring Overview

6.3.2 Dissolved Oxygen, % Saturation Equations O2%Sat = O2Reading/ O2100%Sat Where: O2 reading is the mg/L reading from the RDO Sensor. O2 100%

Page 12 - 5.2 Analog (4-20 mA) 3-wire

© 2008-2013 by In-Situ®, Inc. This document contains proprietary information that is protected by copyright. No part of this document may be photoco

Page 13 - 5.3 SDI-12 (3 wire)

6.4 Calibration Registers Register Size (registers) Mode & Access Level (R/W) Data Type Description 0118 2 R1/W3 float Live salinity value

Page 14

0% Saturation Calibration Values These values represent the sensor conditions while the probe is in a 0% saturation calibration environment. The

Page 15 - 5.5.1 Converter

12. Write the Calibration Mode Off command (0xE002) to the sensor command register to place the sensor in normal operation. If the calibration m

Page 16 - 6.0 Modbus Registers

7.0 Specifications Optical RDO PRO Dissolved Oxygen Probe Sensor type Optical (luminescent) dissolved oxygen sensor RDO PRO Probe Range: 0

Page 17

8.0 Accessories and Replacement Parts Accessory Order No. RDO PRO Sensor Cap Replacement Kit...

Page 18 - Page 18 Modbus Registers

10.0 Appendix A – Communication Device The Communication Device is an accessory product that can be used to calibrate and set up RDO probes. 10.

Page 19 - Page 19 Modbus Registers

11.0 Declaration of Conformity Manufacturer: In-Situ, Inc. 221 East Lincoln Avenue Fort Collins, CO 80524 USA Declares that the followi

Page 20 - 6.4 Calibration Registers

Table of Contents 1.0 Introduction ...

Page 21 - Page 21 Modbus Registers

1.0 Introduction 1.1 System Description The RDO® Pro Probe is a rugged, reliable instrument designed to deliver accurate dissolved oxygen (DO)

Page 22 - Page 22 Modbus Registers

Arrow 1.2 Serial Numbers The instrument serial number is engraved on the side of the unit. The cap serial number is programmed on the memory ch

Page 23 - 7.0 Specifications

2.0 Calibrate the RDO PRO Probe Calibrate the probe with the Comm Kit Software and the Communication Device, or calibrate the sensor directly w

Page 24 - 9.0 Warranty and Service

2.2 2-point Calibration 2.2.1 100% Calibration Point Perform a water-saturated-air calibration as previously described. 2.2.2 0% Calibration

Page 25

3.0 Sensor Deployment The cable end of the RDO PRO Probe is internally threaded (1¼ – 11½ NPT) and can be attached to a male threaded pipe. Whe

Page 26

4.0 Care and Maintenance 4.1 Clean the RDO Cap 1. The cap and nose cone must remain on the probe. 2. Rinse the sensor with clean water from